AddendumIK 5000QUADRA-CHEKSoftware Version3.0.xEnglish (en)12/2013
10 1 Operation1.1 Operation1.1 OperationAdjusting for change of part sizeThe location of features in a part under measure can change due to change i
809541-24 · Ver04 · pdf · 12/2013
IK 5000 QUADRA-CHEK 111.1 OperationPart scaling exampleThis example shows measurement results before and after part scaling has been applied. Start r
12 1 Operation1.1 OperationAuto probe a planePlane measurement accuracy is increased by probing more than the three required points. IK 5000 systems w
IK 5000 QUADRA-CHEK 131.1 OperationDisc pattern auto probeDisc dialogDisc pattern13452Parameter Description1 Center position The center position of th
14 1 Operation1.1 OperationProgramming changesWhile recording a program that is intended to include a looping function (such as a loop, grid, polar gr
IK 5000 QUADRA-CHEK 151.1 OperationProgramming wizardsUse programming wizards to create the following program steps: Grids Polar grids Palletizatio
16 1 Operation1.1 Operation Click the center mouse button to fire the probe tool Click OK in the dialog box
IK 5000 QUADRA-CHEK 171.1 Operation Highlight the first array feature in the Program template as shown Right-click in the Program template and selec
18 1 Operation1.1 Operation Enter the number of features and type in the desired X and Y offsets as shown Click OK in the dialog boxThe grid is crea
IK 5000 QUADRA-CHEK 191.1 OperationPolar gridsA polar grid measures a part with like features that are in a rotational array. For example, create a po
20 1 Operation1.1 Operation Select Measure>Circle Click Enter Point in the Measure Circle dialog Click OK Select the measured Circle in the Pro
IK 5000 QUADRA-CHEK 211.1 OperationPalletizeUse the Palletize wizard for measuring multiple parts placed on a precision fixture.Before palletizing par
22 1 Operation1.1 Operation Right-click in the Program template and select Programming Wizards, then select Palletize Enter the number of features a
IK 5000 QUADRA-CHEK 231.1 Operation Click OK in the dialog boxThe palletization is completed and the program is ready to run. Click New Run to run t
24 1 Operation1.1 OperationRandom placementUse the Random Placement wizard for multiple parts on a stage without a fixture.Before creating a random pl
IK 5000 QUADRA-CHEK 251.1 Operation Click the center mouse button to fire the probe tool Click OK in the dialogThe IK 5000 will use this feature to
26 1 Operation1.1 OperationRandom PlacementTo create a random placement: Highlight the desired features, and the reference point as shown Right-clic
IK 5000 QUADRA-CHEK 271.1 OperationThe IK 5000 inserts the required steps for the random placement.To run a parts program with a random placement: Cl
28 1 Operation1.1 Operation Measure the reference point for the first partRepeat these steps until all parts are entered. Click OK in the dialog box
IK 5000 QUADRA-CHEK 291.1 OperationCopy and paste specialBefore performing a copy and paste special, record a new part program with a datum.To copy an
IK 5000 QUADRA-CHEK 3Information contained in this manualThis addendum covers the operating and setup feature additions and improvements implemented i
30 1 Operation1.1 Operation Click OK in the dialog box Highlight the first array feature in the Program template as shown Right-click in the Progra
IK 5000 QUADRA-CHEK 311.1 Operation Right-click in the Program template and select Paste Special Enter the number of Items and type the desired X an
32 1 Operation1.1 Operation Click OK in the dialog boxThe IK 5000 creates a grid on the part view and assigns a numbered feature name to each feature
IK 5000 QUADRA-CHEK 331.1 OperationThe copied and pasted parts appears in the Part View window as shown.The program is ready to run.
34 1 Operation1.1 OperationAdditional tolerance check Axial Runout for Plane featureAxial runout tolerance can be applied to a Plane feature that is p
IK 5000 QUADRA-CHEK 351.1 OperationPalletizing multiple partsA pallet is a rectangular arrangement of identical parts organized into a consistent matr
36 1 Operation1.1 OperationProfile import pointsProbed points from features can be imported into a profile measurement when recording a program. This
IK 5000 QUADRA-CHEK 371.1 Operation Click Import PtsPoints are imported into the Profile measurement. Click OK
38 1 Operation1.1 OperationMeasuring rectanglesPrecisely five probed points are required to measure a rectangle. The five points must be located in an
IK 5000 QUADRA-CHEK 391.1 OperationProfile3D measurementsProfile3D measurements compare probed part surfaces to nominal part profiles from drawing or
4 Preface
40 1 Operation1.1 OperationProfile3D measurement screen functionsFunctions for conducting Profile3D measurements are located across the bottom of the
IK 5000 QUADRA-CHEK 411.1 OperationProfile3D window menuFunctions for evaluating and displaying Profile3D measurement results are contained in the Mea
42 1 Operation1.1 OperationThe Profile3D measurement processTypical steps required to complete a Profile3D measurement are shown below. Details regard
IK 5000 QUADRA-CHEK 431.1 OperationStart Profile3D measurementTo begin a Profile3D measurement: Click Measure>Profile3Dor Click the Profile3D ico
44 1 Operation1.1 OperationThe nominal Profile3D will be shown in the Measure Profile3D window.Nominal Profile3D shown in Measure Profile3D window
IK 5000 QUADRA-CHEK 451.1 OperationEnable or prohibit data shifts for Profile3D fitThe Profile3D fit algorithm shifts the probed data points in the ca
46 1 Operation1.1 OperationAssign tolerancesBilateral or unilateral tolerances can be assigned to Profile3D measurements. Bilateral tolerances can be
IK 5000 QUADRA-CHEK 471.1 OperationEqual bilateral tolerancesBilateral tolerance values will be centered on the surface of the nominal profile. For ex
48 1 Operation1.1 OperationProbe part surfacesAs part surfaces are probed and points are collected, they appear in green in the Measure Profile3D wind
IK 5000 QUADRA-CHEK 491.1 OperationPerform Profile3D fit analysisWhen all the required points have been probed, the points can be fit to the nominal p
IK 5000 QUADRA-CHEK 51.1 Operation ... 10Adjusting for change of part size ... 10Part scaling example ... 11Auto probe a plane ... 12Grid patt
50 1 Operation1.1 OperationProfile3D automatic fit analysisThe fit analysis can be set to be performed automatically once the measurement is completed
IK 5000 QUADRA-CHEK 511.1 OperationProfile3D Pass Only optionThe Pass Only option on the Profile3D Measure window specifies that the fitting algorithm
52 1 Operation1.1 OperationSix point alignmentA six point alignment allows the operator to align the model with the orientation of the part on the sta
IK 5000 QUADRA-CHEK 531.1 OperationResetting the fit analysisThe results of a fit analysis can be reset and the point data refined in preparation for
54 1 Operation1.1 OperationAdjust the display magnification of error whiskersDuring the fit analysis, probed points that do not fall precisely on the
IK 5000 QUADRA-CHEK 551.1 OperationAdd Profile3D feature to the Feature templateWhen the final fit analysis is finished, the Profile3D measurement can
56 1 Operation1.1 OperationTo view the Profile3D measurement results: Click on the Feature Stamp Icon in the results windowMeasurements results are a
IK 5000 QUADRA-CHEK 571.1 OperationAuto Focus (option)The Auto Focus teach function has been removed from the IK 5000 software. The focus search dista
58 1 Operation1.1 OperationTo set Auto Focus options: Click Tools>Options Verify the Supervisor Password is entered and the Keep privileges until
IK 5000 QUADRA-CHEK 591.1 OperationAuto focus will search and make as many passes as necessary until a optimum focus point is found. The Max two passe
6Auto Focus (option) ... 57Autofocus final check ... 60Exploding the Data Cloud ... 61Rotating and positioning of VED tools using part reference
60 1 Operation1.1 OperationAutofocus final checkAn option to perform a final autofocus check is available from the Auto Focus tab in the Probe Library
IK 5000 QUADRA-CHEK 611.1 OperationExploding the Data CloudExplode Data Cloud is used to extract point features from the data cloud used in a probed f
62 1 Operation1.1 OperationRotating and positioning of VED tools using part reference frameIf a part datum has been performed, the system will rotate
IK 5000 QUADRA-CHEK 631.1 OperationOnce the ISO tolerance is selected the ISO Tolerance Entry dialog box appears. The ISO Tolerance Entry dialog box i
64 1 Operation1.1 OperationThe nominal value for the feature parameter along with the selected ISO tolerance table and index entries are used to deter
IK 5000 QUADRA-CHEK 651.1 OperationAllTol template
66 1 Operation1.1 OperationContinuous probe firingProbes can be dry-fired continuously to show data points that would be acquired under the current li
IK 5000 QUADRA-CHEK 671.1 OperationProgramming continuous probe firingContinuous probe firing is programmable. A Continuous Fire mode step is recorded
68 1 Operation1.1 OperationProgram stepWhen the state of Continuous Fire is changed from the Live Video window or the VED toolbar, a separate Continuo
IK 5000 QUADRA-CHEK 691.1 OperationMeasure shortcutsKeyboard shortcuts for starting Ellipse, Torus, Profile, and Profile3D measurements are now availa
IK 5000 QUADRA-CHEK 72.1 Setup ... 78Updating the IK 5000 software ... 78Backup system settings ... 78Uninstall old software version ... 78Ins
70 1 Operation1.1 OperationSymmetry toleranceA symmetry tolerance can be applied to Line and Plane features that are symmetrical to a reference featur
IK 5000 QUADRA-CHEK 711.1 Operation Apply symmetry tolerance with the constructed line from lines 1 and 2 as the reference featureThe actual toleranc
72 1 Operation1.1 OperationSymmetry tolerance applied to a planeTo apply the symmetry tolerance to a plane: Measure four planes, two planes on opposi
IK 5000 QUADRA-CHEK 731.1 Operation Apply symmetry tolerance with the constructed plane from planes 1 and 2 as the reference featureThe actual tolera
74 1 Operation1.1 OperationPlane probing techniqueA plane can be measured by probing three seed points for the data cloud. The IK 5000 fitting algorit
IK 5000 QUADRA-CHEK 751.1 OperationIn some cases it may be necessary to probe the seed points in a different order and location. For example, probing
76 1 Operation1.1 Operation
Setup
78 2 Setup2.1 Setup2.1 SetupUpdating the IK 5000 softwareThe update should be performed according to the following procedure: Make a backup of the
IK 5000 QUADRA-CHEK 792.1 SetupFiles and foldersIK 5000 files and folders locations are distributed throughout Windows operating system folder locatio
80 2 Setup2.1 SetupPC2-COMP Express frame grabber boardThe IK 5000 supports the PC2-COMP Express, PCI Express frame grabber board produced by Teledyne
IK 5000 QUADRA-CHEK 812.1 SetupUse current reference frameThe Use machine reference option in the Program Properties dialog has been renamed to Use cu
82 2 Setup2.1 SetupThe default state of Use current reference frame is unchecked. If a part reference frame has been created before the program is fir
IK 5000 QUADRA-CHEK 832.1 SetupCamera parametersThe IK 5000 supports the configuration of camera and image quality parameters through DirectShow inter
84 2 Setup2.1 SetupSpecify video probe parametersUse the VED screen in video edge detection systems to specify video probe measurements parameters.To
IK 5000 QUADRA-CHEK 852.1 SetupSpotter camera optionsThe digital camera selected as the spotter camera and the format for the spotter camera can be se
86 2 Setup2.1 Setup Supported image formats: UYVY RGB24 RGB32To access Spotter camera options: Select Tools>Options... Select the VED tab Sel
IK 5000 QUADRA-CHEK 872.1 SetupScale factor for Open Loop joystick control (Optional)A CNC system can be operated in either Closed Loop or Open Loop j
88 2 Setup2.1 SetupSet Scale factor: Click TOOLS>CNC... Enter the Password Click the Joystick tab Enter the desired scaling factor into the Sca
IK 5000 QUADRA-CHEK 892.1 SetupShutdownThe Shutdown function applies to CNC enabled systems only. During system shutdown the probe returns to a positi
Operation
90 2 Setup2.1 SetupDebounce parametersY and Z axis debounce count values for the digital positioner use their own axis value. Previously they used the
IK 5000 QUADRA-CHEK 912.1 SetupTouch probe requalification and starting pointTouch probe auto requalification can be set to Use circular moves and to
92 2 Setup2.1 SetupReset template text colorsIt is possible to set the text color of a template to the same color as the background. When this happens
IK 5000 QUADRA-CHEK 932.1 SetupDefault fixtures files locationA default location for Fixtures files can be selected. The IK 5000 uses this location to
94 2 Setup2.1 SetupExternal amplifier Open Loop control (Optional)IK 5000 Open Loop joystick control provides the ability to toggle between two joysti
IK 5000 QUADRA-CHEK 952.1 SetupJoystick button assignmentSpeed Toggle must be assigned to a joystick button in order use joystick speed control. The H
96 2 Setup2.1 Setup
IndexIK 5000 QUADRA-CHECK 97AAuto focus ... 57Auto probeDisc ... 13Grid .